STELLAR 150 BIPAP- RESMED

ST (Spontaneous–Timed) PAC (Pressure Assisted Control) CPAP (Continuous Positive Airway Pressure) S (Spontaneous) T (Timed)Pressure range IPAP: 2–40 cm H2O EPAP: 2–25 cm H2OTiControl™ Ti Max 0.3–4 sec Ti Min 0.1–Ti MaxRespiratory Rate 5–60 bpmRise Time Min, 150–900 msec (approx.)Trigger and Cycle Five sensitivity settingsAdjustable alarms High Leak Low Minute Ventilation Low and High Pressure Low and High Respiratory Rate Apnoea Low and High FiO2 Low SpO2 Non-vented maskStandard fixed alarms Circuit disconnected Over pressure Blocked tube Internal battery emptyDimensions (LxWxH) 230 x 170 x 120 mmWeight 2.1 kgAir filter Electrostatic fibre meshAir outlet 22 mm taper, compatible with ISO5356–1:2004Internal battery Lithium-ion, 2 hours at EPAP 5, IPAP 15, 20 bpmPower supply AC 100–240V 50–60Hz, 2.2A AC 110V 400Hz, 2.2A DC 24V/3AIEC 60601–1 Classification Class II (double insulation) Type BF Continuous operationData solutions ResScan EasyCare Tx software and Tx Link